Model 287 Ionizer Performance Analyzer The Model 287 performs manual or automatic decay and balance tests for periodic
verification of ionization equipment. It then stores the results and averaged decay times for up to ten manual tests and up to ten complete test sequences.
Temperature and relative humidity are displayed real-time and recorded with the test data. In DECAY mode a built-in high voltage generator charges the plate to
over 1050 volts. During test the plate will discharge toward zero in the presence of ionization. The elapsed time of decay between 1000 volts and 100 volts is
displayed. In the balance mode, isolated plate voltage, test duration and 100 volts is displayed. Self-tests include battery check and tests for functional errors.
|
|
|
Feature 1:
|
All-in-One Instrument -Measures, displays and stores voltage, decay time, temperature and humidity
|
|
Feature 2:
|
Exceeds current requirements of ESD Assn. DSP3.3 for Periodic Verification of Air Ionizers
|
|
Feature 3:
|
Tests balance and decay
|
|
|
|
Specification:
|
Size
|
1.9" H x 4" W x 8.1" L
|
|
Weight
|
1 lb, 6 oz. (0.63 kg)
|
|
Power Requirements
|
9 volt NEDA #1604A or equivalent alkaline
|
|
Meter Range
|
±1250V, 1V resolution
|
|
Plate Capacitance
|
Approximately 25pF
|
|
Drift:
|
<±4V in 90s, ±2V typical
|
|
Trip Points:
|
Fixed 1000V and 100V
|
|
Operating Modes:
|
Manual or Automatic Testing
|
|
Plate Size:
|
1.7"x4" (10.6 inches periphery)
|
|
Accuracy
|
±5% of reading, ±2% typical
|
|
Additional Information:
|
Integrated Temperature and Humidity Sensors:
Humidity ±5% typ. from 10% to 90%RH @25ºC Temperature ±2ºC typical Battery Life- >40 hour life or >1300 charge cycles. Longer life may be
achieved by using 9-volt lithium.
|
|
|
|
Model 287 Ionizer Performance Analyzer Products and Accessories
|
|
|
Description
|
Part Number
|
Unit Price
|
|
|
Model 287 Ionizer Performance Analyzer
|
MME0287
|
$ 995.00
|
ORDER PAGE
|
|
6” x 6” Plate Adapter
|
M287/22C
|
$ 150.00
|
|
Return to Charge Plate Monitors |